You are here: Home

Water test

SIGRIST TurBiScat In-line Process Turbidity & Colour Monitor

p6Applications

  • Filtration monitoring in beverages such as beer fruit juices, spirits and process water
  • Supervision of centrifuges, separators, whirlpools in the beverage industry
  • Turbidity measurement in oils, sugar solutions, food
  • Purity control in chemical and pharmaceutical processes


Industries
  • Beverage
  • Food
  • Chemical Industry
  • Pharmaceutical Industry
Advantages
  • Low-m aintenance design without seals
  • Removable electronic section with quick lock
  • Extended sensor check function with fouling control
  • 2-Angle turbidity measurement with colour compensation
  • Optional colour measurement
  • Easy adjustment with secondary standard
  • Control unit with colour touch screen display
  • Variable display of measuring data, graphs, process performance
  • Smooth system integration using various communication interfaces

pp1Sealless Design
The combination of Hastello sapphire allows operating s unit in practically all process ons – from filtration contro ries up to turbidity control i processes. It greatly reduce for regular maintenance.

 

 

pp2Easy Handling
The electronic section can be quickly removed from the sensor head with a simple grip using a quick lock, no tools are required. Therefore, the sensor head remains in the pipe line. Possible inspection actions can be carried out without interruption of the process.

 

pp3Monitored Safety
An extended sensor check function permanently monitors all system parameters in the background. An optical anti-fouling control provides important information about a successful CIP procedure and informs about necessary cleaning and calibration check. Adjustment check is easily done using a standard.

 

pp4Intelligent Control System
The new control unit SICON uses state-of-the-art touch screen technology with colour display. It allows simple operation using logical menu guidance. Results \displayed can be either plain values, graphs or historical data, including the indication of system and alarm status. The control unit SICON offers all possi-bilities of full system integration using various interface options. A SD card is standard and can be used for data logging with almost unlimited storage capacity for QS requirements.

 

Technical Data

Sensor:

Measuring principle: 90°/25° Scattered light
Wavelength turbidity: LED 650 nm
Wavelength colour (optional): LED 430 nm

Measuring range turbidity:
0 .. 1‘000 EBC
0 .. 4‘000 NTU
0 .. 1‘000 EBC
0 .. 4‘000 NTU
Measuring range colour: 0 .. 50 EBC
Installation: In-line housing Varivent® or compatible
Material sensor head: Stainless steel 1.4301
Windows: Sapphire
Sample temperature: -10 .. +100°C, 180°C with cooling option
Cleaning: CIP/SIP compatible up to 120°C
Pressure: 4 MPa (40 bar) / 100°C
Ambient temperature: -10 .. +50°C
Ambient humidity: 0 .. 100% RH

Protection degree: Wall (pipe) mounting, stainless steel, IP65Panel mounting, aluminium, IP65
Control unit SICON:

Power supply: 9 .. 30 VDC
Power consumption max.: 8 W
Display: 1/4 VGA, 3.5“
Operation: Touch screen
Ambient temperature: -10 .. +50°C
Ambient humidity: 0 .. 100% RH
Protection degree: IP66

Sample connection 6 mm or 1/4” tubing
Ambient temperature -5 to 50 °C (23 to 122 °F)
Humidity 0 to 95% non-condensing relative humidity
Outputs: 4x 0/4 .. 20 mA, galvanic
separated, 7x digital outputs,
5x digital inputs,
freely configurable
Digital Interface: Ethernet, SD-card

Optional: Ethernet, SD-card
Profibus DP, Modbus, Profinet




ClaimEsigrist_logo

 

Main Menu

3100
p3_baner