Water test

Total/Ca-Hardness Analyzer

Model
WATER TEST K-100-TH controller for continuous LOW/HIGHLEVEL Total/Ca-hardness monitoring and control

Applications
On-line monitoring for Calcium (and Magnesium) in boiler water
On-line measurement of Calcium (and Magnesium) after a softening cycle
On-line analysis of Calcium (and Magnesium) in cooling water (e.g. at power plants)
Feedback control of Calcium (and Magnesium) in water (lower/upper set points)With “alarm” output

Methodology
Measurement principle: Direct-potentiometry (Ca-Hardness lon Selective Electrode)with ionic strength adjuster “RISA”) and noise compensated evaluation of sensor data
Unit : “mg/l”
Measuring/control range :
0 – 20 mg/l CaCO3 (0 – 500 mg/l CaCO3 optional)
0.0 – 500 .0 mg/l CaCO
3 (for programmable automatic calibration version)
Detection limit : 1 mg/l CaCO3 (0.1 mg/l programmable automatic calibration version)
Response time (T90) : < 10 min.
Programmable automatic calibration (option)

Advantages
Commonly applied determination method in continuous flow analysis
No sample preparation/filtration (if suspended particles < 1% and < 1 mm)
Easy to set up, minimal maintenance
Environmentally friendly (only non-toxic low-cost reagent NaCl, CaCl2 needed)
Robust full plastic PVC membrane Ca-selective sensor
Analog outputs 0/4 – 20 mA e.g. for curve plotting
Monthly only 1 reagent set (20 l RISA)


Operating principle
P4Via bypass sample water flows through the OVERFLOW CHAMBER where temperature is measured. By action of a peristalticpump a tiny sub-sample is extracted and mixed up with an equal volume of sample conditioner (RISA) in the MIXING TUBE. Thesample/RISA mixture enters into the ELECTRODE HOUSING where the measurement is done; the mixture is finally wasted into a
DRAIN. The temperature and potentimetric signals are transmitted to the controller unit and evaluated. According to the given setpoints (SP1, SP2) reset action/alarm will be triggered (with or without proportionality range).


Specifications

DATA OF CONTROLLER UNIT
- Input: high-impedance. >5x1012 Ohm, galvanically isolated
- Temperature : Pt-100 input, 2-conductor. Range : -30.0… 140.0 +/-0.1 oC
- Temp. compens : manual or automatic
- Display : alphanumerical LCD, 2 lines, 16 characters
- Analog output : 0/4 …20 mA, freely attributable, aglvanically isolated, max.olad 400 Ohm
- Relay output : 2 realy outputs, SPDT contacts 250 V, 6 A max 550 VA
- Interface : optional RS 485
- Ambient temp : operation 0…50 oC storrage -20…65 oC
- Humidity : max 90% at 40 oC non-condensing
- Power required : 230 or 130 or 24 VAC, -10 …+6%, 40…60 Hz
- Power consumption : 10 VA
- Electromagn. Comp : ccording to DIN EN 50081-1 50081-2 in Compliance with CE
- Casing : Polystyrole, spring-loaded terminals
- Protection class : IP 65 (NEMA X4)

DATA OF MEASURING UNIT & HOUSING CABINET
- Sensor : Combined Ca selective matrix membrane electrode with single juction Ag/AgCl reference half cell with 2 ceramic diaphragmas
- Temperature : Separated Pt-100
- Display & range : -1 …20 mg/l CaCO3 (-1 …500 mg/l optional) for manual calibration 0.1-500.0 mg/l CaCO3(for programmable automatic calibration version)
- Precision : +/-2 mg/l (< + 6% for 0 …500 mg/l range) for manual calibration +/- 6% of measuring value or +/- 0.5 mg/l whichever is greater (for programmable automatic calibration)
- Response time : < 10 min
- Waste : Ca-/NaCl
- Maximal pressure : < 1 bar on inlet/pressure-free at outlet of sampling vessel and waste outlet
- Temperature range : 0 …. 40 oC
- Optimal pH range : 3 …9
- Reagent : Reference-cum-Ionic Strength Aduster (RISA) dosing simultaneously over fixed-speed peristaltic pump
- Sample flow rate : 1-6 l/h
- Working life : Sensor electrode : > 1 year
- Cabinet : epoxy coated steel, front door, wall-mounted. IP 55 Class, dimensions (mm) 550 H 450 W 300 L
- Warranty : 1 year

SYSTEM CONFIGURATION
WATER TEST Total/Ca-Hardness controller panel in cabinet mounted. Included:
Overflow chamber, tubes, valve, peristaltic pump, electrode housing, reagent container RISA (20 l), combined Ca-electrode,
Pt 100, controller unit K 100-TH

DISCLAIMER
This publication is not intended to form the basis of a contract and WATER TEST CO., LTD. Reserves, the right to amend

 